General Introduction

Yokogawa OR8ERG is a KCI refillable type ORP sensor with Ryton body which is a strong engineering plastic, which is comparable to PTFE in terms of corrosion resistance and heat resistance, a single type of sensor can support all applications regardless of whether a holder or cleaner is used, integrated-sensor design simplifies calibration with standard solutions and maintenance. It's applicable to work with different holder and converters for wide range of ORP measurement applications.

2.General Specifications

Items

Description

Measuring range

-1500 to 1500 mV

Measuring temperature

-5 to 80°C (refer to table below   when user holder)

Measuring pressure

Atmospheric pressure to   50kPa (refer to table below when user holder)

Temperature   compensation sensor

Pt1000

Wetted part materials

Body:   Ryton (PPS resin), platinum-glass or gold-epoxy resin, titanium, ceramics, FKM

Cable:   Chlorinated polyethylene rubber (Cable sheath)

3.Measurable Gas Components and Ranges

pH Sensor

Holder Type

Holder Material

Cleaner

Temperature

Pressure

OR8ERG

Guide-pipe   PH8HG

PVC

None

-5   to 50°C

Atmospheric   pressure

PP

None

-5   to 80°C

Submersion   PH8HS

Flow-through   PH8HF

PP

None

-5   to 80°C

Atmospheric   pressure (Submersion depth: Max.3 m)

SS

None

-5   to 80°C

Atmospheric   pressure to 10kPa for general purpose

Suspension   HH350G

SS

None

-5   to 80°C

Atmospheric   pressure

Float   PB350G &PB360G

PP,   SS

None

-5   to 50°C

Atmospheric   pressure

  

4.Model Selection

Main Code

Suffix Codes

Description

OR8ERG








KCl Refillable Type ORP Sensor

Electrode

-AU







Gold

-PT







Platinum

Cable Length


-03






3m


-05






5m


-07






7m


-10






10m


-15






15m


-20






20m


-V3






For Variopin connector 3m


-V5






For Variopin connector 5m


-V7






For Variopin connector 10m


-VA






For Variopin connector 7m


-VB






For Variopin connector 15m


-VC






For Variopin connector 20m

Solution Ground Tip



-TN





Titanium

KCl Reserve Tank




-TT1




For general purpose




-TT3




Big volume tank (With 500 mL tank)




-TN1




For maintenance (for TT1)





-N



Always -N

Measuring System






-H


For PH200/PH400






-E


For PH202/FLXA202/FLXA21






-F


For FLXA202/FLXA21






-G


For FLXA402, PH450G,PH202/TB






-V


Variopin connector

Style







*A

Style A

5.External Dimensions

01